PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 ARCTICGEL PADS; ARCTIC SUN GEL PADS Back to Search Results
Catalog Number 317-07
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the pads gave low flow and had to be replaced.
 
Manufacturer Narrative
Received 1 set of 2 unused arcticgel pads.The customer did not return the right thigh pad and the left torso pad for evaluation.Visual evaluation noted evidence of being used, the trim pattern on all four pads was found to be correct, the energy connectors were found free of damages and presented a good connection.Per functional evaluation the pads were submitted to the flow rate test with the arctic sun machine model 2000.The pad was connected to the arctic sun machine model 2000 for 10 minutes and the water immediately started flowing to the pad without any problems.Right chest pad: a total of 5.89 l/min m2 of flow rate were registered during the test.Left thigh pad: a total of 5.09 l/min m2 of flow rate were registered during the test.The reported event was unconfirmed as the reported event could not be reproduced.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use states the following: "once the pads are primed, assure the flow rate displayed on the control panel is greater than 2.3 liters per minute, which is the minimum flow rate for a fullpad kit.¿ (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTICGEL PADS
Type of Device
ARCTIC SUN GEL PADS
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5333149
MDR Text Key35068960
Report Number1018233-2015-00624
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002577
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2016
Device Catalogue Number317-07
Device Lot NumberNGZ10538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-