The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
Received 1 set of 2 unused arcticgel pads.The customer did not return the right thigh pad and the left torso pad for evaluation.Visual evaluation noted evidence of being used, the trim pattern on all four pads was found to be correct, the energy connectors were found free of damages and presented a good connection.Per functional evaluation the pads were submitted to the flow rate test with the arctic sun machine model 2000.The pad was connected to the arctic sun machine model 2000 for 10 minutes and the water immediately started flowing to the pad without any problems.Right chest pad: a total of 5.89 l/min m2 of flow rate were registered during the test.Left thigh pad: a total of 5.09 l/min m2 of flow rate were registered during the test.The reported event was unconfirmed as the reported event could not be reproduced.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use states the following: "once the pads are primed, assure the flow rate displayed on the control panel is greater than 2.3 liters per minute, which is the minimum flow rate for a fullpad kit.¿ (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|