Model Number 3166
Device Problem High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
Device 1 of 4: reference mfr report#1627487-2015-09096; reference mfr report#1627487-2015-09097; reference mfr report#1627487-2015-09098.It was reported the patient did not receive effective stimulation to her pelvic region during previous reprogramming.(event date is unknown).Reportedly, impedance values were high on the right sacral and left thoracic lead contacts.Follow-up identified surgical intervention was undertaken on (b)(6) 2015.During the procedure, 4 leads were explanted and replaced with new models.Effective stimulation was achieved postoperatively.The issue is resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRODE LEAD WIDE SPACED, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston road
plano TX 75024
MDR Report Key5333331
MDR Text Key34639380
Report Number1627487-2015-09094
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2015
Device Model Number3166
Device Lot Number3904370
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-