• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 511.801
Device Problems Device Inoperable (1663); Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the saw attachment device was worn-out and did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was determined that several parts were damaged.The assignable root cause was determined to be due to improper cleaning and maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILLATING SAW ATTACHMENT II
Type of Device
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5334561
MDR Text Key35211900
Report Number8030965-2015-12730
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-