• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a patient experienced an increased intra-peritoneal volume (iipv) event during peritoneal dialysis therapy.It was determined that the patient's last fill was 1000ml and last manual drain was 2400ml.The technical service representative advised the patient to use continuous ambulatory peritoneal dialysis (capd) to complete therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received and an evaluation was performed to investigate the reported event.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.A visual inspection was performed.A short simulated therapy was performed.A review of the event history log did not verify the occurrence of an increased intra-peritoneal volume (iipv) event.Analysis revealed no device malfunction that could have caused or contributed to the reported problem.Upon conclusion of the investigation, the cause of the reported issue was undetermined.Should additional relevant additional information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN  2573
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN   2573
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5335970
MDR Text Key34704309
Report Number1416980-2015-45948
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age74 YR
-
-