Analysis of the instrument and instrument data indicate that the cause for the discordant (b)(6) results is a reagent storage excursion.After repeating samples, a technician stated that the laboratory's reagent refrigerator had changed temperature dramatically to -17c and then went up to room temperature.The technician noticed that some of the reagents were frozen.(b)(6) lot ga5314 was in the laboratory's refrigerator at the time of the temperature fluctuation.That lot was discarded and a new lot shipment, (that was not in that refrigerator until after repairs) was placed on the instrument.No further issues were observed.The instrument is performing within specifications.No further evaluation of the device is required.
|