Catalog Number T5C4441
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received and the evaluation is complete.During the event history log review, an increased intra-peritoneal volume (iipv) event was identified.Upon conclusion of the investigation, the cause of this issue was determined to be use error, the tidal total ultrafiltration removal was set too low.The homechoice apd systems trainer¿s guide gives instructions on how to set the tidal therapy settings.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, one increased intra-peritoneal volume event was identified which occurred in the therapy initiated on (b)(6) 2015 at 23:51:56.During night drain cycle eight, the patient's ultrafiltration reading was 1899ml, indicating the home patient drained 1149ml more than their maximum programmed fill volume of 1500ml.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5338647
MDR Text Key34806823
Report Number1416980-2015-46045
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial
Report Date 12/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberT5C4441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2015
Date Manufacturer Received12/09/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight50
-
-