Catalog Number R5C8320
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, a high drain 101 (night drain #1) alarm was identified in the log.This alarm indicates that the patient drained greater than 200% of the maximum prescribed cycle fill volume.The alarm occurred on (b)(6) 2015 at 14:59:51.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received and the evaluation is complete.During the event history log review, an increased intra-peritoneal volume (iipv) event was identified.Upon conclusion of the investigation, the cause of this issue was undetermined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5338657
MDR Text Key34799001
Report Number1416980-2015-46047
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberR5C8320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2015
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received12/31/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/10/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-