Upon completion of the investigation it was noted that the position of the cam when valve was received was 70mmh2o.The valve was visually inspected: no defects were noted.The valve was irrigated with purified water, no occlusion was noted.The siphon guard was irrigated with purified water, no occlusion was noted.The catheter was irrigated with purified water, no occlusion was noted.The valve was dried.The valve was leak tested, no leaks were noted.The valve was tested for programming.The valve passed the test.The siphon guard was removed.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3832 with lot crcczw, conformed to the specifications when released to stock on the 10th april 2014.Review of the sterilization certificate for the valve conformed to the specification when released on the 4th april 2014.Review of the history device records confirmed the catheters product code 82-3072 with lot crjbn, conformed to the specifications when released to stock on the 11th july 2014.Review of the sterilization certificate for the bactiseal catheter conformed to the specification when released on the 7th july 2014.No root cause could be determined as the valve functions.No problem was found with the bactiseal catheter.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|