Model Number 3660
Device Problem Communication or Transmission Problem (2896)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/21/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient's scs ipg would not pair with the clinician programmer.The programmer(s) would only display an error message when attempting to pair with the ipg.The patient's scs ipg was explanted and replaced which resolved the issue.
 
Manufacturer Narrative
A capa investigation was initiated on 2016-02-23 to investigate the issue in which the ipg experienced loss of power, causing the patient to lose pain relief requiring replacement of the device.Investigation is currently in progress.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5361159
MDR Text Key35703059
Report Number1627487-2016-00171
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/03/2017
Device Model Number3660
Device Lot Number5216875
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-