Model Number 0185 |
Device Problems
Fracture (1260); Pacing Problem (1439); Pocket Stimulation (1463); Ambient Noise Problem (2877); Material Integrity Problem (2978)
|
Patient Problems
Muscle Stimulation (1412); Abdominal Pain (1685)
|
Event Date 10/22/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
|
|
Event Description
|
Boston scientific received information that this patient came into the hospital due to pain/twitching in the stomach.During the follow up many ventricular fibrillation/ventricular tachycardia episodes were seen which weren't true arrhythmias but rather noise resulting in inappropriate anti tachycardia therapy to be delivered.Provocation tests were performed and noise was reproduced.An x-ray was performed which did not reveal a perforation.A revision took place and upon extracting this right ventricular (rv) lead insulation damage as well as a lead fracture was seen.The lead will be returned.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory the complete lead was returned.Visual inspection noted that the trilumen insulation was abraded though the rs- lumen approximately 357 to 373 mm from the terminal pin.There was noted webbing damage between the conductors in the area as well.Microscopic analysis indicated that the damage was initiated by localized compressive stress on the tubing surface.Due to the location and the type of damage it is likely this was caused by entrapment in the clavicle/first rib region.
|
|
Search Alerts/Recalls
|