Model Number G179
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2015
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that during a routine follow-up, all right atrial (ra) lead measurements were appropriate.During the check, the patient began moving in the chair and noise was noted on the ra lead.Atrial impedance measurements during this time were greater than 3,000 ohms.The logbook was reviewed and noted multiple episodes of mode switches due to atrial noise oversensing.As the patient does not require atrial therapy, the entry count to record a mode switch was increased and the system remains implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5372626
MDR Text Key36478502
Report Number2124215-2015-14966
Device Sequence Number1
Product Code OSR
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/14/2016
Device Model NumberG179
Other Device ID NumberAUTOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/10/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
-
-