Boston scientific received information that following the implant procedure, the patient complained of stimulation of the pectoral muscle.Upon interrogation, noise was noted on the pacing and shock channels of the right ventricular (rv) lead.Several episodes were recorded, but no therapy was delivered.During the revision procedure, the rv lead was found to be dislodged.The rv lead was extracted and the helix would not fully extend.As there was tissue in the helix, this rv lead was explanted and replaced.No additional adverse patient effects were reported.
|