Model Number 0292
Device Problems Pocket Stimulation (1463); Capturing Problem (2891); Device Dislodged or Dislocated (2923)
Patient Problem Muscle Stimulation (1412)
Event Date 11/14/2015
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this patient implanted with a right ventricular (rv) lead experienced muscle stimulation.The field representative tried all left ventricular (lv) configurations and turned lv pacing off.Additional information obtained from the field which indicated that the physician thought there was dislodgement of lead which had caused the diaphragmatic stimulation and high pacing threshold measurements.The lead was repositioned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5376282
MDR Text Key36240482
Report Number2124215-2016-00964
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/10/2017
Device Model Number0292
Other Device ID NumberRELIANCE 4-SITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4135
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age35 YR
-
-