DEPUY SYNTHES POWER TOOLS K-WIRE ATTACHM F/EPD+APD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.037 |
Device Problems
Mechanical Problem (1384); Physical Resistance (2578); Mechanical Jam (2983)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper cleaning and maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by the (b)(6) that during service and evaluation, it was observed that the attachment device bearing seized, was jammed, and heavy moving.It was further noted that the grip fingers and bearings of the extension piece are worn.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|