MAUDE Adverse Event Report: FRESENIUS MEDICAL CARE NORTH AMERICA FRESENIUS COMBISET 2008; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS MEDICAL CARE NORTH AMERICA FRESENIUS COMBISET 2008; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problems Fluid/Blood Leak (1250); Misconnection (1399)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not yet been returned for physical evaluation.A plant investigation is still ongoing and a supplemental report will be submitted upon completion.
 
Event Description
An inpatient user facility reported a blood leak about ten minutes into patient treatment.The venous line disconnected at the hub after the lines were adjusted by a patient care technician, which resulted in a blood leak at this location.The line was reconnected, the treatment was terminated and the patient set-up on the same machine with different bloodlines.The patient then continued treatment and was able to complete therapy successfully.No adverse symptoms or other events occurred and no medical intervention was required.Estimated blood loss was 200cc.Bloodlines are available and have been requested for evaluation.
 
Manufacturer Narrative
The device was returned to the manufacturer for physical evaluation.A visual examination of the returned device was performed; no defects were identified.The device was received with an unknown connector (a non-fmc product) attached to the bloodline.A visual examination of the unknown connector revealed the presence of a crack.A leak test was then performed on the complaint device by pressuring the arterial and venous lines and submerging the device in water.During this test, air bubbles were observed coming from the unknown connector.A second leak test was performed with the return device, but during this test, the unknown connector was removed and replaced to with the recirculation connector (a fmc product).No air bubbles were observed and no leaks were identified.A dimensional analysis of the male conical fittings of the venous and arterial patient connectors was performed; the dimensional checks were within the acceptable range.A records review was performed on the reported lot.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.The investigation into the cause of the reported problem was not able to confirm the failure mode.The complaint device was received with an unknown connector (a non-fmc product) attached.Although a leak was observed due to the crack in the unknown connector, when the non-fmc product was switched for the proper recirculation connector, the device functioned as intended; no leak observed.Therefore, this complaint has been deemed not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRESENIUS COMBISET 2008
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
FRESENIUS MEDICAL CARE NORTH AMERICA
reynosa 88780
MX  88780
Manufacturer (Section G)
REYNOSA
erika de reynosa, s. a.
mike allen 1331 parque
reynosa, tamaulipas cp 88780
MX   88780
Manufacturer Contact
tanya taft
920 winter st.
waltham, MA 02451
7816999000
MDR Report Key5396918
MDR Text Key37237530
Report Number8030665-2016-00039
Device Sequence Number1
Product Code FJK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Medical Technologist
Type of Report Initial,Followup
Report Date 03/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2018
Device Catalogue Number03-2722-9
Device Lot Number15DR01207
Other Device ID Number00840861100293
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2016
Initial Date FDA Received01/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient Weight81
-
-