Catalog Number 293001-001
Device Problems Low Battery (2584); Battery Problem (2885); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reported that the electrical contacts of the companion external battery are pushed inside the housing.The customer also reported that the hospital staff subsequently exchanged the companion external battery.There was no reported adverse patient impact.This alleged failure mode poses a low risk to the patient since the low battery did not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver has a redundant, alternate power source of external wall power.The companion external battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reported that the electrical contacts of the companion external battery were pushed inside the housing.The customer also reported that the hospital staff subsequently exchanged the companion external battery.There was no reported adverse patient impact.The companion external battery was returned to syncardia for evaluation.Visual inspection revealed that the j1 and j2 connectors were completely recessed into the external battery's housing, rendering the external battery inoperative.The investigation could not determine how the connectors became recessed, but it was likely the result of an impact shock.The companion external battery was taken out of service.This failure mode posed a low risk to the patient because it would not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver has redundant, alternate power sources of external wall power and an internal, emergency battery.Customers are also provided with several backup external batteries.This is the first reported instance of this issue.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK
Type of Device
BATTERY PACK
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5397045
MDR Text Key37680494
Report Number3003761017-2016-00028
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 01/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number293001-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/21/2016
Initial Date FDA Received01/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age47 YR
-
-