STRYKER ENDOSCOPY-SAN JOSE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
|
Back to Search Results |
|
Catalog Number 0250080617 |
Device Problems
Break (1069); Crack (1135); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The following repair diagnostic codes were identified: cracked/peeling insulation at tip of shaft.This does confirm the alleged failure mode of [insulation scan failed].Probable root cause: poor autoclave reliability.Incorrect sterilization/reprocessing procedure.Handling procedures.Contact forces.Product used beyond defined useful life.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that insulation cracked, compromised, broke, or breached (failed insulscan).
|
|