Catalog Number 0250080234 |
Device Problems
Crack (1135); Component Missing (2306)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported third party insulation and there was missing parts on handle and there is a cracked rotation knob.
|
|
Manufacturer Narrative
|
The following repair and service diagnostic codes were identified: broken/fractured handle; replaced handle.The following was observed: broken / fractured handle.Handle was replaced.This does confirm the alleged failure mode of: insulation damage.Probable root cause: "poor autoclave reliability; incorrect sterilization/reprocessing procedure; handling procedures; contact forces; product used beyond defined useful life." the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported third party insulation and there was missing parts on handle and there is a cracked rotation knob.
|
|
Search Alerts/Recalls
|