Model Number 3660
Device Problems Device Inoperable (1663); Communication or Transmission Problem (2896)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/23/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported no communication could be established with the patient's ipg.An error message was displayed.A magnet was used to attempt to establish communication to no avail.As a result, surgical intervention is planned for a later date to address the issue.
 
Event Description
Additional information received identified surgical intervention took place on (b)(6) 2016 at which time the patient's ipg was explanted and replaced.Effective stimulation was reportedly restored postoperative.
 
Manufacturer Narrative
The capa was initiated on 2016-02-23 to address the issue in which the ipg experienced loss of power, causing the patient to loose pain relief requiring replacement of the device.Investigation is currently in progress.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.Corrected data: results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5414265
MDR Text Key37610486
Report Number1627487-2016-00568
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup,Followup
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/24/2017
Device Model Number3660
Device Lot Number5286200
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/08/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/15/2016
Initial Date FDA Received02/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
06/02/2017
Supplement Dates FDA Received02/22/2016
09/21/2016
06/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-060217-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age37 YR
-
-