CAREFUSION ALARIS EXTENSION SET; FILTER, INFUSION LINE Back to Search Results
Model Number 20029E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
The customer reported a leaking filter extension set while on a patient.No patient harm reported.
 
Manufacturer Narrative
The customer¿s report of a leak was not confirmed.Visual inspection observed no obvious damages or any issues.Functional testing was performed.The set was attempted to be reprimed via a lab syringe filled with blue dyed fluid.No leaks or any issues were observed anywhere on the set.A lab extension set was also attached to the set's luer end and was pressure tested up to 30psi while submerged underwater in which no leaks or any issues were also observed.The root cause of the reported leak was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS EXTENSION SET
Type of Device
FILTER, INFUSION LINE
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
stephen bilello
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5416076
MDR Text Key37720311
Report Number9616066-2016-00232
Device Sequence Number1
Product Code FPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K801614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 01/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20029E
Device Catalogue Number20029E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2016
Initial Date FDA Received02/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received05/20/2016
06/15/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-