• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problem Moisture Damage (1405)
Patient Problem No Information (3190)
Event Date 08/22/2013
Event Type  malfunction  
Manufacturer Narrative
This initial report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.(b)(4).Based on the reported event, this is a known issue that was addressed by an internal action.Provided that the faulty component is returned, at this time another investigation does not need to be performed.
 
Event Description
The customer reported a lot of water was found inside the pressure transducer.Additionally the map was at 5cmh2o without any gas source turned on.Although a lot of water was blown out of the transducer, customer requested replacement.No additional information was received.Patient involvement is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
7149227830
MDR Report Key5416344
MDR Text Key37730438
Report Number2021710-2016-02967
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/22/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3100B
Device Catalogue Number773967
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/22/2013
Date Device Manufactured12/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-