• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO CONSULTA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO CONSULTA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D224TRK
Device Problems Premature Discharge of Battery (1057); Output above Specifications (1432)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/12/2015
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary - the device was returned and analyzed.Analysis of the device revealed normal battery depletion.(b)(4).
 
Event Description
It was reported that the device had early eri (elective replacement indicator).It was noted that chronic higher left ventricular (lv) output may have contributed to higher current drain.The device was explanted and replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULTA CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5423940
MDR Text Key37849388
Report Number3004209178-2016-02043
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/14/2012
Device Model NumberD224TRK
Device Catalogue NumberD224TRK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
407652 LEAD, 694765 LEAD, 419588 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00089 YR
-
-