MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems Loss of Power (1475); Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Date 12/05/2014
Event Type  malfunction  
Manufacturer Narrative
This medwatch report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.The customer was sent a replacement unit.The unit was returned to carefusion for evaluation and the issue was isolated to the alarm board pn 768588.
 
Event Description
The following description of the event was documented by a carefusion tech support specialist in a response to a phone conversation with a user facility representative on (b)(6) 2014.The customer called to report that this unit completely lost power on two separate occasions on a patient.It did alarm correctly.He states that the rt's were at the bedside and were able to respond right away and turn it back on with no compromise to the patient.He said they ended up swapping the unit out with a different vent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5427328
MDR Text Key38535799
Report Number2021710-2016-03113
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-