The customer reported that they received questionable results for approximately six patient samples tested for tina-quant hemoglobin a1c gen.2 (hba1c) on a c111 analyzer.The customer loaded a new hba1c reagent kit on the analyzer and initially tested the patient samples.Patient values were said to be either too high or too low.The samples were repeated after the reagent kit was replaced and calibrated.Of the six patient samples, two were found to have erroneous hba1c results.It was stated that none of the erroneous results were reported to the patient.It was not clear if any erroneous results were reported outside of the laboratory.A clarification has been requested.The first sample initially resulted as 4.8%.After the reagent kit was replaced, the sample was repeated, resulting as 6.69%.The second sample initially resulted as 7.0%.The sample was repeated outside of the laboratory on a different analyzer, resulting as 9.0%.After the reagent kit was replaced on the original analyzer, the sample was repeated again, resulting as 8.9%.The patients were not adversely affected.The c111 analyzer serial number was (b)(4).It was stated that the customer was not running quality controls and that they were doing weekly interlab comparisons with hba1c hplc method.The customer stated that they did not have any further issues after replacing the reagent cassette.
|