A risk to health could not be excluded for these specific circumstances, although: no injury of a patient or user or any other person has occurred (and no patient in the treatment room at the time of event).There is no indication of a systematic error or malfunction of the brainlab device.There is no indication that the brainlab device would not have worked as intended.Corresponding measures to minimize this anticipated risk as low as reasonably practicable are already in place.According to the results of the brainlab investigation and the information provided by the hospital, it can be concluded that the root cause for the m3 falling down is that an inadequate and insufficient attachment procedure of the m3 micro-multileaf collimator was performed in this specific case.The mounting-procedure as described in the corresponding user guide was not followed; specifically the device was not correctly locked into position, and the status of the relevant gantry accessory mount interlock lights was not reviewed.Therefore the removable m3 micro-multileaf collimator was not properly mounted and was still able to slide out.Investigation by brainlab (on-site and off-site) has shown that the fixation mechanism of this m3 worked as intended.Even with the m3 damaged after the event, the m3 could still be mounted to the accessory mount with all locking mechanisms engaging.There is no indication of an error or malfunction of the device, the device worked as intended.Corresponding measures to minimize this anticipated risk as low as reasonably practicable are already in place.The m3 micro-multileaf collimator has been returned from this site to brainlab after the event for repair.Despite according to the hospital the operators at this site are informed about the correct procedure to mount the m3 on the linear accelerator, brainlab intends to offer an additional training to this hospital.
|