Merge hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.Customer reported that on the live waveform display that the ffr is showing.78 for the recording at 12:42:42am however in the recorded waveform he cannot get that value of.78 from any of the recorded beats.This was found to be due to the software design.The system is designed to calculate ffr two different ways on the client pc (for chron log recording) and on the hemomonitor for display.On the hemomonitor the system calculates ffr by taking the quotient of the average of pa and the average of pd, and displays that minimum value.In the client pc, the system takes the average of the quotient of pa and pd.If there is a difference between the value displayed on the hemomonitor and the value recorded in the chron log, and the cardiologist uses the value displayed on the hemomonitor for clinical decisions, there is a potential of over treatment or undertreatment.(b)(4).
|