The reported incorrect product packaged could not be verified as product associated with this event has not been provided for review.A review of the provided radiographs was inconclusive due to poor resolution quality.The device history record review, the non-conformance history review, and the etq complaint history review did not provide an indication of a manufacturing deviation.The review of scrap part labels and supporting production logs for a consecutive run of int3211 and int411 was completed for a potential lot mix-up, and no anomalies were identified.
|