Model Number MSERIES BI-PHASIC
Device Problem Failure to Charge (1085)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to zoll medical corporation; the reported malfunction was observed and attributed to high contact resistance within the front panel membrane.The front panel membrane was replaced to remedy the problem.Analysis for reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during biomed testing, the device failed to charge.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5439005
MDR Text Key38668448
Report Number1220908-2016-00316
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 01/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Other Device ID Number00847946000684
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/27/2016
Initial Date FDA Received02/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-