• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 01/25/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Field advisory: 1627487-07262012-001-c.This ipg serial number was included in a field advisory and a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2.Reference mfr.Report# 1627487-2016-00725.It was reported the patient ((b)(6)) experienced pocket heating at the ipg site while charging the ipg for about 45 minutes.A sjm representative met with the patient and determined it was the user issue.The issue was resolved and the patient decided to keep the original charger.Date of implant is unknown.On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Event Description
Device 1 of 2 : reference mfr.Report# 1627487-2016-00725.Further follow up identified the implant date for the device, is updated with an availbale information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5446843
MDR Text Key38554567
Report Number1627487-2016-00724
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2013
Device Model Number3788
Device Lot Number3286456
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/18/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-