The customer reported there was over-sensing and over pacing when the tpm lead was inserted in the patient and connected to the external pulse generator (epg).Later there was undersensing of the qrs and pacing on t-wave, causing the patient to go into life threatening ventricular arrhythmia.The patient required cardioversion.Due to the problems reported, the facility checked the cables and identified two of the cables were problematic.During troubleshooting, the customer found that when the cable was moved after switching on the epg the sensing light flickerd even without external connection.The epg seemed to be functioning properly until the cables were moved.
|