• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC. MEDTRONIC REUSABLE EXTENSION CABLE; ELECTRODE CABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. MEDTRONIC REUSABLE EXTENSION CABLE; ELECTRODE CABLE Back to Search Results
Model Number 53912
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/06/2015
Event Type  malfunction  
Manufacturer Narrative
The device was in use for treatment.A review of the device history record could not be performed as the lot number of the cable is unknown.The cable in question was returned for analysis.A visual inspection found the device returned with the black connector detached from the cable and the strain relief removed.The resistance through both cables was found to be within specification; however, the resistance measurement for the solid grey cable did not include the black connector.The cable failed at the solder joint.The instructions for use (ifu) informs the user that the cable can be re-sterilized by oscor eto gas sterilization a maximum of two times.Precautions are provided to the user: do not connect any cable model to a/c power source; connection of exposed pins to a/c power source may pose a risk of serious injury or death.Additional information regarding this event was requested but not provided.A follow-up report will be sent if additional information becomes available.A corrective and preventive action has been opened to address this issue.
 
Event Description
The customer reported that the cable is breaking.There was no report of any patient adverse effects related to this event.
 
Manufacturer Narrative
(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC REUSABLE EXTENSION CABLE
Type of Device
ELECTRODE CABLE
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer Contact
jan flegeau
3816 desoto blvd.
palm harbor, FL 34683-1618
7279372511
MDR Report Key5449834
MDR Text Key38645255
Report Number1035166-2016-00018
Device Sequence Number1
Product Code IKD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070926
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number53912
Device Catalogue Number53912
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/24/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-