Catalog Number 10697306 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/21/2016 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
The root cause of the discordant sodium, calcium and potassium results is unknown.Siemens is in process of investigating the event.
|
|
Event Description
|
Customer reported discordant sodium, potassium and calcium results on the instrument.There was no report of injury due to this event.
|
|
Manufacturer Narrative
|
Siemens field service engineer (fse) evaluated customer affected device.The valve on sn (b)(4) was found to be out of the alignment specification of ± 0.003" on 3/4/16.The cartridge interface frame was replaced by field service and verified in specification.Results appear typical since replacement although benzalkonium interference is observed on several occurrences on this instrument.Benzalkonium is a known interferent on the rp500 na sensor per the rp500 operators manual.
|
|