SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500 Back to Search Results
Catalog Number 10697306
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2016
Event Type  No Answer Provided  
Manufacturer Narrative
The root cause of the discordant sodium, calcium and potassium results is unknown.Siemens is in process of investigating the event.
 
Event Description
Customer reported discordant sodium, potassium and calcium results on the instrument.There was no report of injury due to this event.
 
Manufacturer Narrative
Siemens field service engineer (fse) evaluated customer affected device.The valve on sn (b)(4) was found to be out of the alignment specification of ± 0.003" on 3/4/16.The cartridge interface frame was replaced by field service and verified in specification.Results appear typical since replacement although benzalkonium interference is observed on several occurrences on this instrument.Benzalkonium is a known interferent on the rp500 na sensor per the rp500 operators manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5450738
MDR Text Key39209592
Report Number1217157-2016-00019
Device Sequence Number1
Product Code KHP
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10697306
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-