The reported issues involve the following syncardia temporary total artificial heart (tah-t) system components and are reported under two separate medical device reports: primary freedom driver s/n (b)(4) (mfr report # 3003761017-2016-00063 and backup freedom driver s/n (b)(4) (mfr report # 3003761017-2016-00064).The customer reported that the primary freedom driver s/n (b)(4) exhibited a fault alarm while the patient was switching out his onboard batteries.The customer also reported that the patient was subsequently switched to the backup freedom driver s/n (b)(4).The customer also reported that the patient noticed the power adaptor port on freedom driver s/n (b)(4) is broken.The customer also reported that the patient was subsequently switched to another backup freedom driver.There was no reported adverse patient impact.The freedom driver s/n (b)(4) was returned to syncardia for evaluation.Visual inspection of the exterior of the driver revealed a broken connection 1 receptacle, which connects the power adaptor to the driver, confirming the customer-reported issue.Visual inspection of the interior of the driver revealed no abnormalities.Despite the broken connection 1 receptacle, the driver in as received condition met all pressure test acceptance criteria, which included normotensive and hypertensive settings, with no anomalies or alarms.The root cause of the broken connection 1 receptacle reported by the customer and verified during visual inspection, is consistent with the result of an impact shock to the driver.Despite the customer-reported issue, risk to the patient was low because it did not prevent the driver from performing its life-sustaining functions.The freedom driver is also equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and an external battery charger which provide additional means of mitigating the impact to patient care in the event that the power adapter cannot be connected.Freedom driver s/n (b)(6) was serviced.Based on days of use, the service included of the main printed circuit board assembly (pcba), speaker pcba, piston cylinder assembly (pca), and motor controller pcba, before being placed into finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|