MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number RPM 20-320
Device Problems Break (1069); Material Twisted/Bent (2981); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 01/20/2016
Event Type  malfunction  
Manufacturer Narrative
Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(6).A supplemental medwatch will be submitted, if additional information becomes available.
 
Event Description
It was reported that during priming "noisy rpm, broken and twisted belts, repair done under work order # (b)(4) by replacing belt kits" no patient involvement.(b)(4).
 
Manufacturer Narrative
A maquet service technician evaluated the device and replaced the defective parts.The device was then returned to the customer for use.Although the parts were returned to the manufacturing site for further evaluation, the device was not.Manufacturing were therefore unable to determine the root cause of the issue.No further issues have been reported by the customer since the parts were replaced.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5455870
MDR Text Key39269148
Report Number8010762-2016-00109
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRPM 20-320
Device Catalogue Number70102.7652
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2016
Initial Date FDA Received02/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse