MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number RPM 20-320 |
Device Problems
Break (1069); Material Twisted/Bent (2981); Noise, Audible (3273)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/20/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(6).A supplemental medwatch will be submitted, if additional information becomes available.
|
|
Event Description
|
It was reported that during priming "noisy rpm, broken and twisted belts, repair done under work order # (b)(4) by replacing belt kits" no patient involvement.(b)(4).
|
|
Manufacturer Narrative
|
A maquet service technician evaluated the device and replaced the defective parts.The device was then returned to the customer for use.Although the parts were returned to the manufacturing site for further evaluation, the device was not.Manufacturing were therefore unable to determine the root cause of the issue.No further issues have been reported by the customer since the parts were replaced.
|
|
Event Description
|
(b)(4).
|
|