• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 5MM PEEK MULTIFUNCTION HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 5MM PEEK MULTIFUNCTION HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250181106
Device Problems Break (1069); Crack (1135); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/02/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported the insulation cracked, compromised, broke, or breached.Insulscan failed.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Dents and scratches were seen on the insulation.The instrument was tested for cracks in which an insul scan test was performed and failed.The probable root causes are normal wear, improper sterilization methods, and user misuse.(b)(4).
 
Event Description
It was reported the insulation cracked, compromised, broke, or breached.Insulscan failed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM PEEK MULTIFUNCTION HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5458685
MDR Text Key39304108
Report Number0002936485-2016-00191
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250181106
Device Lot Number¿1447989H¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/02/2016
Initial Date FDA Received02/24/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-