• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635TU20060
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported that a valve leak occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman® access system (was) was positioned in the laa.The physician was not able to tighten the hemostatic valve completely, allowing for additional back bleed.The was exchanged for another and the procedure was completed successfully.There were no patient complications reported and the patient's status is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN® ACCESS SYSTEM
Type of Device
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5460389
MDR Text Key39016909
Report Number2134265-2016-01515
Device Sequence Number1
Product Code NGV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 01/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberM635TU20060
Device Catalogue NumberTU2006
Device Lot Number18643074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number91033431 FA
Patient Sequence Number1
-
-