• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (CRM-KISTA) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (CRM-KISTA) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5256
Device Problems Difficult to Interrogate (1331); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a follow up in (b)(6) 2015, the pulse generator exhibited loss of telemetry.At follow up on (b)(6) 2016 the device exhibited intermittent telemetry and could not be fully interrogated despite using two different programmers.The magnet rate was approximately 90 ppm.The device was explanted and replaced.The patient was in good condition before, during and post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITY ADX XL
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key5470459
MDR Text Key39357165
Report Number3010215456-2016-00667
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2005
Device Model Number5256
Device Lot Number2993491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2016
Initial Date FDA Received03/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-