MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. KIT 640G BLACK MMOL; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. KIT 640G BLACK MMOL; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-1751W
Device Problems No Display/Image (1183); Out-Of-Box Failure (2311)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/10/2016
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that new insulin pump out of the box was damaged.It was reported that after battery was installed and insulin pump was turned on, display screen was blank and grayed out.Customer's blood glucose was not reported.
 
Manufacturer Narrative
Device powered up properly after battery installation.No blank or gray display noted.Device was monitored for two (2) days and no blank display, gray display or additional display anomalies noted.Device received with scratched case and pillowing keypad overlay.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT 640G BLACK MMOL
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5470800
MDR Text Key39643528
Report Number3004209178-2016-89752
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1751W
Device Catalogue NumberMMT-1751W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-