• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problem Charging Problem (2892)
Patient Problems Inadequate Pain Relief (2388); Weight Changes (2607)
Event Date 12/31/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
The patient reported stimulatiion was lost approximately 2 weeks ago and the charger cannot locate the ipg.Reportedly, a replacement charger was sent to no avail.The patient programmer displays an error message.In addition, space was added between the paddle and ipg due to the patient previously losing 50lbs.The patient further states the ipg started to move around in the pocket.As a result, surgical intervention was decided as the next course of action.Follow-up idenified the ipg was explanted and replaced during surgery on (b)(6) 2016.
 
Event Description
Follow-up identified the issue is resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5472348
MDR Text Key39415284
Report Number1627487-2016-01016
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup,Followup
Report Date 04/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/28/2016
Device Model Number3788
Device Lot Number4444731
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2016
Initial Date Manufacturer Received 02/09/2016
Initial Date FDA Received03/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received03/14/2016
03/18/2016
04/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age54 YR
-
-