MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiopulmonary Arrest (1765); Death (1802)
Event Date 02/08/2016
Event Type  Death  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient expired due to cardiopulmonary arrest.There is no known allegation from a health professional that the death was related to the device.No further information is available.
 
Manufacturer Narrative
Analysis of the device noted high voltage circuitry damage in the laboratory.Review of the device image confirmed that the high voltage circuitry was normal on the date of the reported event of death.The high voltage circuit damage occurred after the death date.The device was tested on the bench as well as using automated test equipment and the low voltage circuitry was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5483338
MDR Text Key39766877
Report Number2938836-2016-01627
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Model NumberCD3365-40Q
Device Lot Number4189961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2016
Initial Date FDA Received03/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
Patient Outcome(s) Death;