• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS MEDICAL SOLUTIONS USA INC. ACUSION ; DIAGNOSTIC ULTRASOUND DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS MEDICAL SOLUTIONS USA INC. ACUSION ; DIAGNOSTIC ULTRASOUND DEVICE Back to Search Results
Model Number ACUSON SC2000
Device Problem Computer Software Problem (1112)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/14/2014
Event Type  malfunction  
Manufacturer Narrative
The issue is in non thread safe access to a list of continuous capture uids when creating the list ( a null dataset is added to the list).Later when stress echo asks for the list, tho parser throws a null reference exception and nothing is sent to stress echo task f(an empty list) so stress echo doesn't know about continuous capture clips.The issue was fixed in a software release.
 
Event Description
Loss of post exercise images taken during some stress echo exams were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSION
Type of Device
DIAGNOSTIC ULTRASOUND DEVICE
Manufacturer (Section D)
SIEMENS MEDICAL SOLUTIONS USA INC.
685 east middlefield road
mountain view CA 94043
Manufacturer Contact
karen smith, vp-regulatory affa
685 east middlefield road
mountain view, CA 94043
8004228766
MDR Report Key5485619
MDR Text Key39938234
Report Number3009498591-2015-00069
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
072786
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Not Applicable
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 01/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACUSON SC2000
Device Catalogue Number10433816
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/15/2014
Initial Date FDA Received06/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-