• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM10120
Device Problems Break (1069); Difficult or Delayed Positioning (1157); Fracture (1260); Malposition of Device (2616); Activation, Positioning or Separation Problem (2906)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2016
Event Type  Injury  
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide further patient details.
 
Event Description
It was reported that during deployment of the endovascular stent graft in a dialysis case, the outer catheter snapped.The stent graft could be deployed with much effort by using hemostats.Due to the deployment difficulties, the stent graft was not deployed in the correct position at the target site and therefore, a second stent graft was used to cover the lesion completely.No patient injury was reported.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.On the basis of the condition of the returned device, the reported deployment issue could be confirmed.The outer catheter was found to be elongated indicating the presence of high deployment force.The system was found to be fractured distal to the handle.Reportedly, the stent graft could be successfully implanted.As no images were provided, the reported malposition of the stent graft could not be verified.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.The event may be associated with difficult anatomic conditions leading to increased friction during deployment and subsequent sheath fracture.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." and "examine the packaging and endovascular system to determine if there is any damage, defects or if the sterile barrier is compromised.Do not use the device if any of these conditions are observed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5488435
MDR Text Key39934761
Report Number9681442-2016-00071
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008622
UDI-Public(01)04049519008622(17)180612(10)ANZF1436
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029/S003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/12/2018
Device Catalogue NumberFEM10120
Device Lot NumberANZF1436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2016
Initial Date FDA Received03/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR
Patient Weight75
-
-