Model Number 3788
Device Problem Communication or Transmission Problem (2896)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the sjm representative met with the patient and determined the ipg is unable to communicate with multiple external devices.The patient has been without stimulation for three months.The patient was recharging for one hour per week and has been unable to recharge despite numerous attempts.Surgical intervention may be pending to address this issue.Date of event is unknown.
 
Event Description
Follow up information identified the patient underwent surgical intervention to remove and replace the ipg which resolved the issue.Therapy has been restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5490197
MDR Text Key40015979
Report Number1627487-2016-01202
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/28/2014
Device Model Number3788
Device Lot Number3719245
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/06/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-