A steris field service technician arrived onsite, inspected the surgical lights, and confirmed the reported damage.The technician applied electrical tape to the damaged wires, tested the lights, and confirmed them to be operating according to specification.Steris's inspection of the wires evidenced damage to the insulation surrounding the wires.Over time, various manipulations of the lighthead assembly wore the damaged wire down and allowed for the wire to become exposed.The amsco examiner 10 surgical light complies with ul 544 and iec 60601-1 requirements.The surgical light is not under steris contract agreement for maintenance.The hospital's biomedical technician is responsible for maintenance.
|