• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM10120
Device Problems Break (1069); Positioning Failure (1158); Fracture (1260); Premature Activation (1484)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/05/2016
Event Type  Injury  
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any patient details.
 
Event Description
It was reported that during placement at the apex of a forearm graft, the endovascular stent graft partially deployed resulting in the breakage of the outer sheath of the delivery system at the y-adapter.The stenting procedure was stopped at that time.No patient injury was reported.This is the same patient as reported in medwatch report # 9681442-2016-00074.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The evaluation of the returned device confirmed the reported event.The stent graft was found to partially deployed.The outer sheath was found to be elongated and torn off indicating that increased friction must have affected on the delivery system during the attempt to deploy the stent graft.Due to the fracture of the outer sheath, a complete stent graft deployment was impossible.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent sheath fracture.In this case, no information regarding the anatomy of the tracking vessel and the target site was provided.Insufficient flushing of the device may be another contributing factor.In this case, no information regarding the preparation of the device was provided.On the basis of the information available, a definite root cause for the reported failure could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5492925
MDR Text Key40118462
Report Number9681442-2016-00073
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008622
UDI-Public(01)04049519008622(17)180529(10)ANZE2711
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029/S003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/29/2018
Device Catalogue NumberFEM10120
Device Lot NumberANZE2711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-