• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, MARLBORUGH MA MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, MARLBORUGH MA MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM Back to Search Results
Catalog Number 10-401
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/11/2016
Event Type  malfunction  
Manufacturer Narrative
Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Currently unable to establish a relationship or impact to the reported observation.Myosure disposable: according to the instructions for use (ifu) precautions: exercise care when inserting or removing the device.Excessive bending of the device distal tip can cause the trd's cutter to come out of the cutting window.If such damage occurs, replace the device immediately.(b)(4).
 
Event Description
It was reported during a myosure procedure for uterine tissue removal on (b)(6) 2016, the "window broke off in the patient in 2 pieces".The physician then removed the pieces without incident and completed the procedure with another disposable device.No injury to the patient.The patient was discharged home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Type of Device
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, MARLBORUGH MA
250 campus drive
marlborough MA 01752
Manufacturer Contact
callahan
250 campus drive
marlborough, MA 01752
5082638859
MDR Report Key5493337
MDR Text Key40326445
Report Number1222780-2016-00054
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/03/2018
Device Catalogue Number10-401
Device Lot Number15L03RJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-