The freedom power adaptor was not supporting a patient.The customer reported that the freedom power adaptor has a broken mounting rail.The freedom power adaptor was returned to syncardia for evaluation.Visual inspection of the freedom power adaptor revealed damage to the mounting rail and indicator light lens.The freedom power adaptor was functionally tested and passed all functional testing.Because of the damage to the mounting rail and indicator light lens, the freedom power adaptor was taken out of service.This failure mode poses a low risk to a patient because the issue was observed when the freedom power adaptor was not supporting a patient.In addition, it would not prevent the driver from performing its life-sustaining functions.
|