Model Number 3788 |
Device Problems
Nonstandard Device (1420); Communication or Transmission Problem (2896)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).This ipg serial number was included in multiple field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
The patient reported the she has not been receiving stimulation for the past six months and the ipg has been unable to communicate with the charging system.Surgical intervention may be pending to address this issue.Date of event is unknown.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
Follow up information identified the patient underwent surgical intervention on (b)(6) 2016 where the ipg was removed and replaced.The ipg was inoperable due to lack of recharging.Therapy has been resumed and issue has been resolved.
|
|