• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/TEE,TBG & MTHPC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/TEE,TBG & MTHPC Back to Search Results
Catalog Number 1883
Device Problems Inaccurate Delivery (2339); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.No photo for review.A device history record review could not be conducted since the lot number was not provided.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.However material from the production line was inspected and no issues were detected that can lead to this customer complaint.If the device sample becomes available at a later date this complaint will be re-opened.
 
Event Description
The customer alleges that the micro mist nebulizer is not nebulizing the medication completely (at 8 to 10 lpm).There's too much residual medication in the cup post treatment.No patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON MICRO MIST NEBULIZER W/TEE,TBG & MTHPC
Type of Device
NEBULIZER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
margie burton, rn
3015 carrington mill blvd
morrisville, NC 27560
9194334965
MDR Report Key5503999
MDR Text Key40430246
Report Number3004365956-2016-00156
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Respiratory Therapist
Type of Report Initial
Report Date 03/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number1883
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-