Model Number 3716 |
Device Problem
No Device Output (1435)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 02/23/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.Udi (di): (b)(4).
|
|
Event Description
|
It was reported the patient's ipg is inoperable.Subsequently, the patient may undergo surgical intervention as the next course of action.
|
|
Event Description
|
Follow-up information revealed the sjm representative met with the patient and it was found that the patient's programmer locates the implant and displays a " ipg low battery "stim off" warning.In addition, the patient's 3711 charging system does not locate the ipg.A replacement charging system was sent to the patient as a troubleshooting step.
|
|
Event Description
|
Follow-up information revealed the patient underwent surgical intervention where her ipg was explanted and replaced.The patient reported effective stimulation therapy postoperative.
|
|